Boonton 4540 Series RF Power Meter

Operation

4-11

4.5.7 Channel 1 | 2 Menu

The Channel menu contains items that affect a single measurement channel. The ―channel‖ is a full measurement path,
beginning with an external power sensor and sensor cable, and including data acquisition, processing of measurements, and
display of the processed information. The Model 4541 single channel power meter has only a Channel 1 menu, while the
dual-channel Model 4542 has separate color-coded menus for Channel 1 and Channel 2.

Main > Channel > Channel 1 | 2 > On/Off

Description:

Set or return the measurement state of the selected
channel.

When

ON,

the

channel

performs

measurements; when OFF, the channel is disabled and
no measurements are performed.


Menu Type:

Toggle.

SCPI Command:

CALCulate[1|2]:STATe <Boolean>


Main > Channel > Channel 1 | 2 > Vert Scale

Description:

Set the power or voltage vertical sensitivity of the trace
display in channel units.


Menu Type:

Numeric Value.


Range:

Watts: 1 nW/Div to 50MW/Div

Volts: 1 mV/Div to 50kV/Div

dBm,

dBV,

dBr:

0.1 dB/Div to 50 dB/Div

SCPI Command:

DISPlay:TRACe[1|2]:VSCALe <numeric_value>

Main > Channel > Channel 1 | 2 > Vert Center

Description:

Set or return the power or voltage level of the
horizontal centerline of the graph for the specified
channel in channel units. If a change in the vertical
scale causes the center maximum value to be exceeded, the center will be forced to the
maximum value for the new range.

Menu Type:

Numeric Value.


Range:

Watts,

Volts: ±10,000 * Vertical Scale per divsion

dBm,

dBV,

dBr:

±200 dB


SCPI Command:

DISPlay:TRACe[1|2]:VCENTer <numeric_value>





Fig. 4-7 Channel 1|2 Menu