Advanced features and settings – Boonton PIM 21 User Manual User Manual

Page 19

Advertising
background image

19

Boonton

5. Antennas are intentional radiators and can couple energy into other PIM generating

structures in close proximity. Depending on the level of PIM generation within the
adjacent structure the PIM can, in turn, also be received by the antenna and then be
measured by the Test Set. In this case it is difficult to know if it is the antenna at fault
or if the PIM is due to the environment around it.

Advanced Features

and Settings

The PIM 21 has a number of features and adjustable parameters to allow for customization
and differing test requirements.

The MENU key is used to scroll between 3 different menu items;

DISPLAY

SETUP

EXIT

After scrolling with the MENU key the desired topic is selected with the SELECT key. Selecting
escape will take you back to the previous menu.

A quick escape back to the default RF OFF display is to hold down the MENU key and press the
SELECT key until the display cycles through to RF OFF.

The UP DOWN keys are only used to increment the setting from within the SETUP menu.

Menu Structure
Start with default RF OFF display, then press MENU to reach either DISPLAY or SETUP:

From DISPLAY press SELECT, then press MENU to cycle through the following items:

1. Rx RSSI, Displays receiver sensitivity in dBm, typically -126 dBm*
2. PIM, Displays the units of measurement of PIM, in dBc
3. F1 Fwd PA Pwr, Displays the units of measurement for power of F1 as 0 dBm
4. F2 Fwd PA Pwr, Displays the units of measurement for power of F2 as 0 dBm
5. F1 Return Loss, Displays the units of measurement of F1 return loss as 0 dBm
6. F2 Return Loss, Displays the units of measurement of F2 return loss as 0 dBm
7. Battery Volts, Displays internal battery volts, 12 Volts*
8. Input Volts, (Ext DC) Displays battery volts being applied to EXT DC connector
9. 4.0 Volts (internal rail) Displays internal 4 Volt Rail, 4 Volts*
10. 5.0 Volts (internal rail) Displays internal 5 Volt Rail, 5 Volts*

EXIT, to return to default RF OFF hold MENU button down and press the SELECT key until the
display cycles through to RF OFF.

*If measurement deviates from this approximate value, call for service.