Table 42, Inputs and outputs specifications (both models) – Vaisala DMT346 User Manual

Page 164

Advertising
background image

User's Guide _______________________________________________________________________

164 __________________________________________________________________ M210762EN-G

Table 42

Inputs and Outputs Specifications (Both Models)

Property

Value

Operating voltage

10 ... 35 VDC, 24 VAC

with optional power supply module

100 ... 240 VAC 50/60 Hz

Default start-up time

initial reading after power-up

3 s

full operation after sensor Purge and
AutoCal

Approximately 6 min

Power consumption at 20 °C
(U

in

24 VDC)

U

out

2 × 0 ... 1V / 0 ... 5V / 0 ... 10V

Max 25 mA

I

out

2 × 0 ... 20 mA

Max 60 mA

RS-232

Max 25 mA

Display and backlight

+ 20 mA

During sensor purge and warming

+ 110 mA max

Analog outputs (2 standard, 3rd
optional)

Current output

0 ... 20 mA, 4 ... 20 mA

Voltage output

0 ... 1 V, 0 ... 5 V, 0 ... 10 V

Accuracy of analog outputs at 20 °C

± 0.05 % full scale

Temperature dependence of the analog
outputs

± 0.005 %/°C full scale

External loads

Current outputs

R

L

< 500

Ω

0 ... 1V output

R

L

> 2 k

Ω

0 ... 5V and 0 ... 10V outputs

R

L

> 10 k

Ω

Wire size

0.5 ... 2.5 mm

2

(AWG 20 ... 14)

stranded wires recommended

Digital outputs

RS-232
RS-422/485 (optional)
LAN (optional)
WLAN (optional)

Protocols

ASCII commands
Modbus RTU
Modbus TCP

Relay outputs (optional)

0.5 A, 250 VAC, SPDT

Display (optional)

LCD with backlight, graphic trend
display

Menu languages

Chinese, English, Finnish, French,
German, Japanese, Russian,
Spanish, Swedish