Figure 7, Ball valve installation – Vaisala MM70 User Manual
Page 20
USER'S GUIDE____________________________________________________________________
3.
Make sure that the temperature at the measurement point is
equal to that of the process, otherwise the moisture reading may
be incorrect.
0605-085
Figure 7
Ball Valve Installation
The following numbes refer to Figure 7 on page 18 above.
1 = Probe
2 = Clasp nut. The probe slides easily in place when you first fix
the clasp nut manually. Then complete the installation and
make it stable by tightening the clasp nut with a plastic
wrench. Do not overtighten the screw.
3 = Handle of the ball valve
4 = Process
chamber/pipeline
5 = Manual press tool
6 = The groove on the probe indicates the upper adjustment limit
7 = Ball of the ball valve
18 __________________________________________________________________ M210878EN-A