Aalborg SMV Series User Manual

Page 13

Advertising
background image

5.2 Troubleshooting Guide

For best results it is recommended that instruments are returned to the fac-
tory for servicing. See section 1.3 for return procedures.

5.3 Technical

Assistance

Aalborg

®

Instruments will provide technical assistance over the phone to qual-

ified repair personnel. Please call our Technical Assistance at 800-866-3837.
Please have your Serial Number and Model Number ready when you call.

9

INDICATION

LIKELY REASON

REMEDY

no response to
"speed" and "direction"
control signals

inadequate gas
pressure

apply appropriate gas
pressure

cable or connector
malfunction

check cables and all
connections or replace

valve out of adjustment return to factory for

repair / replacement

valve does not work in
OPEN position

incorrect valve
adjustment

return to factory for
repair / replacement

pc board defect

return to factory for
repair / replacement

cable or connectors
malfunction

check cable and
connectors or replace

valve does not work in
CLOSE position

incorrect valve
adjustment

return to factory for
repair / replacement

pc board defect

return to factory for
repair / replacement

cable or connectors
malfunction

check cable and
connectors or replace