Aalborg PSV-D User Manual

Page 11

Advertising
background image

8

INDICATION

LIKELY REASON

REMEDY

valve does
not work in
“active" position

incorrect valve adjustment

re adjust valve

pc board defect

return to factory for replacement

cable or connector malfunction

check cable and connectors or
replace

differential pressure too high

decrease pressure to correct level

insufficient inlet pressure

adjust appropriately

valve does not
work in close
position

incorrect valve adjustment

re adjust valve

pc board defect

return to factory for replacement

cable or connector malfunction

check cable and connectors or
replace

orifice obstructed

disassemble to remove impedi-
ments or return to factory

For best results it is recommended that instruments are returned to the factory
for servicing. See section a.3 for return procedures.

f.3 Technical

Assistance

Aalborg Instruments will provide technical assistance over the phone to qualified
repair personnel. Please call our Technical Assistance at 845-770-3000. Please
have your Serial Number and Model Number ready when you call.