Overview, Specifications – Campbell Scientific CS215 Temperature and Relative Humidity Probe User Manual

Page 14

Advertising
background image

CS215 Temperature and Relative Humidity Probe

8. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the

PC to datalogger connection is active, you can click

Finish in Short Cut

and you will be prompted to send the program just created to the

datalogger.

9. If the sensor is connected to the datalogger, as shown in the wiring

diagram in step 6, check the output of the sensor in the datalogger support

software data display to make sure it is making reasonable measurements.

5. Overview

The CS215 probe uses a single chip element that incorporates both a

temperature and an RH sensor. Each element is individually calibrated with

the calibration corrections stored on the chip. The element is easily changed in

the field. The ability to replace the element in the field reduces downtime and

calibration costs.

Electronics within the CS215 control the measurement made by the sensor

element, apply temperature and linearization corrections to the readings, and

present the data via SDI-12 to a datalogger.

A sintered plastic filter minimizes the effects of dust and dirt on the sensor.

The filter is lightweight and hydrophobic, thereby diminishing its effect on the

time response of the sensor.

The probe housing is designed to withstand permanent exposure to all weather

and to fit into a range of radiation shields, including compact shields.

6. Specifications

Features:

Accurate, stable measurements

Field-changeable element allows on-site recalibration

Individually calibrated sensor elements require no further adjustment

of the probe

Low power consumption

Digital SDI-12 output

Compatible with Campbell Scientific CRBasic dataloggers:

CR200(X) series, CR800 series, CR1000, CR3000, and CR5000.

Also compatible with Edlog dataloggers: CR500, CR510, CR10(X),

and CR23X

Sensor Element:

Sensirion SHT75

Supply Voltage:

6 to 16 Vdc

Current Consumption: Typically 120 µA quiescent

1.7 mA during measurement

Diameter:

12 mm at sensor tip, maximum 18 mm at cable end

Length:

180 mm, including cable strain relief

Housing Material:

Anodized aluminum

4

Advertising