7.3 Installation

Locate the sensor over an open, level area at least 9 m (EPA) in diameter. The

surface should be covered by short grass or the natural earth surface where

grass does not grow. Sensors should be located at a distance of at least four

times the height of any nearby obstruction and at least 30 m (EPA) from large

paved areas. Sensors should be protected from thermal radiation and

adequately ventilated. Protect the filter at the top of the sensor from exposure

to liquid water. The hydrophobic nature of the filter repels light rain, but

driving rain can force itself into the pore structure of the filter and take time to

dry out.

Standard measurement heights:

1.5 m (AASC)

1.25 – 2.0 m (WMO)

2.0 m (EPA)

See Section 10, Attributions and References, for a list of references that discuss

temperature and relative humidity sensors.

When used in the field, the CS215 must be housed in a radiation shield.

Typically, the 41303-5A six-plate solar radiation shield is used. The white

color reflects solar radiation, and the louvered construction allows air to pass

freely through, thereby keeping the probe at or near ambient temperature.

The 41303-5A attaches to a crossarm, mast, or user-supplied pipe with a 2.5 to

5.3 cm (1.0 to 2.1 inch) outer diameter.

A 41003-5 ten-plate solar radiation shield can also house the CS215 sensor by

using the 18.5 adapter (pn 6637). For this shield, tighten the sensor grip

around the probe where it best matches the size of the grip. (The probe will

also fit directly into most of the R.M. Young shields, where the probe enters

the shield from the top, pointing downwards.)

Both the 41303-5A and the 41003-5 solar radiation shields attach to a

crossarm, mast, or user-supplied pipe with a 2.5 to 5.3 cm (1.0 to 2.1 inch)

outer diameter.

Tools required for installing a radiation shield to a tripod or tower include:

1/2-inch open-end wrench

small screwdriver provided with datalogger

small Phillips screwdriver

UV-resistant cable ties

small pair of diagonal-cutting pliers

1. Loosen the plastic split collar at the base of the shield (reversing the

removable portion if necessary) and gently insert the probe.

2. Tighten the collar until it lightly grips the probe body.

3. Continue to push the probe up into the body of the shield until the step in

the tube stops it from going any further.