Campbell Scientific SR50-L CSC Ultrasonic Distance Sensor User Manual

Page 15

Advertising
background image

SR50 Sonic Ranging Sensor

9

TABLE 1. SR50 SDI-12 Command List

CR10X and

SDI-12

FUNCTION

DATA OUTPUT

CR23X P105

COMMANDS

Parameter 2

When using

Options non-Campbell
Loggers

0

aM

First Target (Meters)

D1

1

aM1

First Target (Meters)

D1,Q1

with

Quality

Number

2

aM2

Multiple Targets (Meters)

D1,D2,D3

3

aM3

Multiple Targets (Meters)

D1,D2,D3,Q1,Q2,Q3

with

Quality

Numbers

4

aM4

First Target (Feet)

D1

5

aM5

First Target (Feet)

D1,Q1

with Quality Number

6

aM6

Multiple Targets (Feet)

D1,D2,D3

7

aM7

Multiple Targets (Feet)

D1,D2,D3,Q1,Q2,Q3

with

Quality

Numbers

8

aM8

First Target (Meters)

D1

No Retries

N/A aD0 Send

Data

DEPENDENT UPON

COMMAND

SENT

10

aV

Start Verification calculates

NUMERIC SIGNATURE

ROM

Signature

aI

Send Identification

012CAMPBELLSR50 1.1

NOTE: There are 2 spaces between SR50 and 1.1

0-- aC Concurrent

Measurement*

D1

n-- aCn

Concurrent Measurements*

Same as M – M8

Where a = address of SDI-12 device.
Where n = numbers 1 to 8.

* = see APPENDIX A for details about Concurrent Measurements.

Advertising