Altek a 830-KP User Manual

Page 10

Advertising
background image

17

OPERATING INSTRUCTIONS

Instrument with

RTD Input
Controller

Temperature Transmitter

Temperature Indicator

Temperature Trip or Alarm

Receiver Input

Ohms or 2-Wire RTD

Controller

Transmitter

Computer

Three Wire RTD Connection

Two Wire RTD Connection

CALIBRATE RTD INPUTS

Four Wire RTD Connection

SENSOR

}

+ COMP

- COMP

Jumper

Jumper

16

CALIBRATE RTD INPUTS

Choose this function to simulate a temperature signal into any
instrument requiring an RTD input.

1) Disconnect the RTD from the instrument being calibrated.
2) Press the T/C/RTD/

pushbutton until any RTD type (Pt, Ni or

Cu) appears on the display.

3) Press the DISPLAY/SOURCE/READ push-button until SOURCE

and any RTD TYPE appear on the display

4) Connect using 2 or 3 wires as in the diagrams on the opposite

page. Spade lugs are recommended to minimize any contact
resistance.
Output temperature is continuously adjustable with the
UP/DOWN push-buttons. Zero & Span (or any other values)
are available by using the LO and HI "QUIK-CHEKs”.

To Change the RTD type
1) Press the TYPE/ENG UNITS push-button. The words SELECT

RTD TYPE and the current selection appear on the display.

2) Press any UP/DOWN push-button until the required type

appears on the display.

3) Press the STORE/RESET push-button or wait 5 seconds to

store the selection

Note on 4-Wire Connections:

The TechChek 830-KP may be used to calibrate instruments
requiring a 4-Wire RTD connection. This connection is valid
only when the lead wires included with the 830-KP are used to
connect the 830-KP directly to the instrument. Use of longer
wires can cause errors in the output setting.

For long 4-Wire cable runs use the 3-Wire connection and add
a stacking banana jack (not available from Transmation) to
connect a second wire to jack #1. Connect this extra wire to the
fourth wire field connection.

OPERATING INSTRUCTIONS