SPECIFICATIONS

SOURCE RTD & OHMS

RTD TYPES:

Pt 100

for 1.3850 (DIN/IEC 751 & New JIS), 1.3902 (Burns),

1.3926 (US Lab) & 1.3916 (Old JIS 1604C-1981)
Ni 120

, Cu 10

& Cu 50

RTD RESOLUTION: 1°C or 1°F
RANGE OHMS: 0.0 to 400.0 Ohms
ACCURACY: ±0.05% of Full Scale + 0.075 mV/mA Excitation Current
ACCURACY OHMS: ±(0.05% of 400.0 Ohms + 0.1 Ohm = ±0.3 Ohms
(At 1 mA Excitation Current)
TEMPERATURE EFFECT: ±((0.035 mV/°C)*(1/mA Excitation Current))
ALLOWABLE EXCITATION CURRENT: 0.125 to 2.0 mA continuous DC

READ RTD

RTD TYPES & RESOLUTION: Same as for SOURCE RTD
RTD RANGE (IN OHMS): 0.0 to 400.0 Ohms
RTD ACCURACY (OHMS): ±(0.05% of 400.0 Ohms + 0.1 Ohm = ±0.3 Ohms
EXCITATION CURRENT SUPPLIED: 1 mA, nominal

FREQUENCY SOURCE

RANGES: 1 to 1500 CPM (Count-Per-Minute); 1 to 1500 Hz, 0.01 to 15.00 kHz
ACCURACY:

±(0.05% of 1500 CPM + 1 CPM) = ±2 CPM
±(0.05% of 1500 Hz + 1 Hz) = ±2 Hz
±(0.05% of 15.00 kHz + 0.01 kHz) = ±0.02 kHz

OUTPUT WAVEFORM: Square Wave, Zero Crossing, -1V to +5V ±10%
RISETIME: Hz <5 microseconds; CPM <100 microseconds
OUTPUT IMPEDANCE: <100 Ohms
SOURCE CURRENT: >1 mA at 10 kHz
SHORT CIRCUIT DURATION: Infinite

MEASURE FREQUENCY

RANGES & ACCURACY: Same as for FREQUENCY SOURCE
TRIGGER LEVEL: 1 V RMS, DC coupled to 10.25 V;

7 V RMS, DC coupled to 250 V

INPUT IMPEDANCE: > 1Meg Ohm + 60 pF

38

SPECIFICATIONS

MEASURE DC VOLTS

RANGES:

0.00 to 110.00 mV; 0.00 to 10.25 V; 0.0 to 200.0 V

ACCURACY:

±(0.05% of 110 mV+ 0.01mV) = ±0.07 mV
±(0.05% of 10.25 V + 0.01V) = ±0.02V
±(2% of 200.0 V + 0.1V) = ±4.1 V

INPUT RESISTANCE: >1 Meg Ohm to 10.25V, >5 Meg Ohm to 200V
SOURCE RESISTANCE EFFECT: 0.01% per 100 Ohms

SOURCE THERMOCOUPLES

THERMOCOUPLE TYPES: J, K, T, E, N, R & S
RESOLUTION: 1°C or 1°F
ACCURACY: °C ±(0.05% OF 80 MV + 1°C); °F ±(0.05% OF 80 MV + 1°F)
COLD JUNCTION ACCURACY: ±1°C
COLD JUNCTION EFFECT: within 0.05°C per °C change
OUTPUT IMPEDANCE: <0.3 Ohms
SOURCE CURRENT: >10 mA, Max

READ THERMOCOUPLES

THERMOCOUPLE TYPES & ACCURACY: Same as for SOURCE T/C
RESOLUTION: 0.1°C or 0.1°F
COLD JUNCTION ACCURACY: ±1°C
COLD JUNCTION EFFECT: within 0.05°C per °C change
INPUT IMPEDANCE: > 1 Meg Ohm
OPEN THERMOCOUPLE DETECTION: 450 millisecond pulse.
Nominal threshold, 10 K Ohms.

READ OHMS

RANGE OHMS: 0.0 to 1000.0 Ohms
ACCURACY: ±(0.05% of 1000.0 Ohms + 0.1 Ohm) = ±0.6 Ohms
EXCITATION CURRENT SUPPLIED: 1 mA, nominal

CONTINUITY CHECKING

TEST CURRENT: Nominal 1 mA
THRESHOLD: 100 Ohm ±20%
INDICATION: Steady tone & Symbol on LCD plus Ohm Reading