Altek a 830-KP User Manual

Page 15

Advertising
background image

27

BLACK

BLACK

RED

3

2

1

RTD Element

OPERATING INSTRUCTIONS

MEASURE RTD SENSORS

26

MEASURE RTD SENSORS

Choose this function to read an RTD. Three wires must be use for
both 2 and three wire RTDs.

1) Disconnect the RTD from the instrument being calibrated.
2) Press the T/C/RTD/

pushbutton until any RTD type (Pt, Ni or

Cu) appears on the display.

3) Press the DISPLAY/SOURCE/READ push-button until SOURCE

and any RTD TYPE appear on the display

4) Connect using 3 wires as in the diagrams on the opposite

page. Use a 3 wire connection to read a 4 wire RTD (all four
wires must be the same gauge). Spade lugs are recommend-
ed to minimize any contact resistance.

To Change the RTD type
1) Press the TYPE/ENG UNITS push-button. The words SELECT

RTD TYPE and the current selection appear on the display.

2) Press any UP/DOWN push-button until the required type

appears on the display.

3) Press the STORE/RESET push-button or wait 5 seconds to

store the selection.

OPERATING INSTRUCTIONS

Advertising