InstaCal Error Messages

If an InstaCal module that has not been characterized in more than one year is connected to
the Site Master, the follow message will be displayed:

The InstaCal Module Connected, S/N XXXXX
Was Characterized more than 1 year ago.
Characterization date
Current SiteMaster Date
You are advised to return this InstaCal module
to the factory for re-characterization.
Press ENTER to continue, see manual for details.

Return the InstaCal module to the factory for re-characterization in order to ensure contin-
ued accurate measurements.

If the serial number of the connected InstaCal module does not match the serial number
stored in the Site Master, the following message is displayed:

The InstaCal characterization data stored in the SiteMaster
is for a module different than the one currently connected.
SiteMaster contains data for InstaCal module S/N:
Currently connected InstaCal Module S/N:
Would you like to overwrite the previously
loaded InstaCal characterization?

Press the

YES

soft key to update the stored InstaCal characterization to use the currently

connected module.

Press the

NO

soft key to keep the stored InstaCal characterization.

2-22

Chapter 2 Functions and Operations

Error Message

Description

CANNOT ZERO
NO DETECTOR
INSTALLED

Attempting to perform a Power Monitor zero adjust function with no
RF detector connected to the Site Master.

CANNOT ZERO
INPUT SIGNAL
TOO HIGH

Attempting to perform a Power Monitor zero adjust function with an
input of greater than –20 dBm.

POWER MONITOR
OPTION NOT
INSTALLED

Attempting to enter Power Monitor mode with no Option 5 installed.

Table 2-3.

Range Error Messages (2 of 2)