The following keypad hard key functions are printed in blue on the keypad keys.

Turns the liquid crystal display (LCD) back-lighting ON or OFF.
(Leaving back lighting off conserves battery power.)

LCD Contrast adjust. Use the Up/Down arrow key and

ENTER

to adjust

the display contrast.

AUTO

SCALE

Automatically scales the display for optimum resolution.

LIMIT

Displays the limit line menu for the current operating mode.

MARKER

Displays the marker menu of the current operating mode.

PRINT

Prints the current display to the selected printer via the RS232 serial port.

RECALL

DISPLAY

Recalls a previously saved trace from memory. When the key is pressed, a
Recall Trace selection box appears on the display. Select a trace using the
Up/Down arrow key and press the

ENTER

key to implement.

To erase a saved trace, highlight the trace and select the

DELETE TRACE

softkey. To erase all saved traces, select the

DELETE ALL TRACES

softkey.

RECALL

SETUP

Recalls a previously saved setup from memory location 1 through 10.
When the key is pressed, a

RECALL SETUP

selection box appears on the

display. Select a setup using the Up/Down arrow key and press the

ENTER

key to implement. Setup 0 recalls the factory preset state.

RUN

HOLD

When in the Hold mode, this key starts the Site Master sweeping and pro-
vides a Single Sweep Mode trigger; when in the Run mode, it pauses the
sweep. When in the Hold mode, the hold symbol (page 2-19) appears on
the display. Hold mode conserves battery power.

SAVE

DISPLAY

Saves up to 200 displayed traces to non-volatile memory. When the key is
pressed,

TRACE NAME:

appears in the lower left of the display. Save the

display with up to 16 alphanumeric characters for that trace name and
press the

ENTER

key to implement.

SAVE

SETUP

Saves the current system setup to 1 of 10 internal non-volatile memory lo-
cations. When the key is pressed, a

SAVE SETUP

selection box appears

on the display. Use the Up/Down arrow key to select a setup and press the

ENTER

key to implement.

START

CAL

Starts the calibration in SWR, Return Loss, Cable Loss, or DTF measure-
ment modes (not available in Spectrum Analyzer mode).

2-5