Range er rors 2-21 – Anritsu S331C User Manual

Page 35

Advertising
background image

2-20

Chapter 2 Functions and Operations

Error Message

Description

BATTERY LOW

Battery voltage is less than 9.5 volts. Charge battery. If condition per-
sists, call your Anritsu Service Center.

EXTERNAL
POWER LOW

External supply voltage is less than 12.5 volts. Call your Anritsu Ser-
vice Center

PLL FAILED

Phase-locked loops failed to lock. Charge battery. If condition persists
with a fully charged battery, call your Anritsu Service Center

INTEGRATOR
FAILED

Integration circuit could not charge to a valid level. Charge battery. If
condition persists with a fully charged battery, call your Anritsu Ser-
vice Center
.

EEPROM R/W
FAILED

Non-volatile memory system has failed. Call your Anritsu Service
Center.

OUT OF TEMP.
RANGE

Ambient temperature is not within the specified operating range. If the
temperature is within the specified operating range and the condition
persists, call your Anritsu Service Center.

RTC BATTERY
LOW

The internal real-time clock battery is low. A low or drained clock bat-
tery will affect the date stamp on saved traces. Contact your nearest
Anritsu Service Center.

LO LOCK FAIL

The local oscillator in the spectrum analyzer has phase lock loop er-
rors. If condition persists with a fully charged battery, call your Anritsu
Service Center
.

POWER MONITOR
FAIL

Failure of the power monitor system board voltages. If condition per-
sists, call your Anritsu Service Center.

BATTERY CAL
LOST

Battery communication failed. The indicated battery charge status may
be invalid. If condition persists, call your Anritsu Service Center.

MEMORY FAIL

The EEPROM test on the Site Master main board has failed. If condi-
tion persists, call your Anritsu Service Center.

The time and date
Have not been set
on this Site Master.
To set it, after exit-
ing here press
the <SYS>
[CLOCK] keys.
Press ENTER or
ESC to continue

The time and date are not properly set in the Site Master. If condition
persists, call your Anritsu Service Center.

Note: A listing of Anritsu Service Centers is provided in Table 1-2 , page 1-8.

Table 2-2.

Self Test Error Messages

Advertising
This manual is related to the following products: