Instacal mod ule ver i fi ca tion 3-5, Manual calibration procedure, Chapter 3 getting started – Anritsu S331C User Manual

Page 45: Start cal, Enter, Figure 3-2. calibration setup

Advertising
background image

Manual Calibration Procedure

If the “CAL OFF” message is displayed, or the test port cable has been changed, a new cali-
bration is required. The following procedure details how to perform the calibration.

Step 1. Select the appropriate frequency range, as described in the procedure above.

Step 2. Press the

START CAL

key. The message “CONNECT OPEN or InstaCal TO RF

Out PORT” will appear in the display.

Step 3. Connect the calibrated Open and press the

ENTER

key. The messages “Mea-

suring OPEN” and “CONNECT SHORT TO RF Out” will appear.

Step 4. Remove the Open, connect the calibrated Short and press the

ENTER

key. The

messages “Measuring SHORT” and “CONNECT LOAD TO RF Out” will ap-
pear.

Step 5. Remove the Short, connect the calibrated Termination and press the

ENTER

key. The messages “Measuring LOAD” will appear.

Step 6. Verify that the calibration has been properly performed by checking that the

“CAL ON” message is now displayed in the upper left corner of the display.

3-4

Chapter 3 Getting Started

HOLD

RUN

START
CAL

AUTO
SCALE

SAVE
SETUP

RECALL
SETUP

LIMIT

MARKER

SAVE
DISPLAY

RECALL
DISPLAY

PRINT

MODE

FREQ/DIST

AMPLITUDE

SWEEP

SYS

ENTER

CLEAR

ESCAPE

ON

OFF

/

1

2

4

5

6

7

8

9

0

3

+

-

.

Site Master S332C

OPEN

LOAD

SHORT

CALIBRATION

RFOUT/REFLECTION

TEST PORT

TEST PORT CABLE (OPTIONAL)

Figure 3-2.

Calibration Setup

Advertising
This manual is related to the following products: