Setting the Scale

Auto Scale

The Site Master can automatically set the scales to the minimum and maximum values of
the measurement on the y-axis of the display. This function is particularly useful for mea-
surements in SWR mode. To automatically set the scales:

Step 1. Press the

AUTO SCALE

key.

The Site Master will automatically set the top and bottom scales to the minimum
and maximum values of the measurement on the y-axis of the display.

Amplitude Scale

The following procedure sets the top and bottom scale display.

Step 2. Press the

AMPLITUDE

key to call up the Scale Menu.

Step 3. Press the

TOP

soft key and use the keypad or Up/Down arrow key to edit the top

scale value.

Step 4. Press the

BOTTOM

soft key and use the keypad or Up/Down arrow key to edit

the bottom scale value.

NOTE: Typically the y-axis scale of the display is 0-54 dB (return loss) but for
some measurements (for example, insertion loss) the scale should be changed
to 0-10 dB. If the scale is not changed, some measurement results may not be
easily displayed on the screen.

Save and Recall a Setup

Saving a Setup

Saving a setup configuration in memory will preserve the calibration information.

Step 1. To save the configuration in one of the 10 available user setup locations, press

SAVE SETUP .

Step 2. Use the key pad or the Up/Down arrow key to select a location (1 - 10).

Step 3. Press

ENTER

to save the setup.

NOTE: A manual calibration setup will be saved with an OSL designator. An
InstaCal setup will be saved with an ICAL designation.

3-7

Chapter 3 Getting Started