Step 3. Press the

MODE

key.

Step 4. Use the Up/Down arrow key to select the Spectrum Analyzer mode and press

ENTER

.

Step 5. Select the appropriate reference level, input attenuation level, resolution band-

width, and video bandwidth.

Step 6. Press the

MEAS

key.

Step 7. Once in the MEAS menu, press the

ACP

soft key.

Step 8. Select the center frequency by pressing the

CENTER FREQ

soft key and enter

the desired center frequency.

Step 9. Select the main channel bandwidth by pressing the

MAIN CHANNEL BW

soft

key and enter the desired main channel bandwidth. For an IS-95 signal enter
1.23 MHz.

Step 10. Select the adjacent channel bandwidth by pressing the

ADJ CHANNEL BW

soft

key and enter the desired adjacent channel bandwidth. For an IS-95 signal enter
30 kHz.

Step 11. Select the channel spacing by pressing the

CHANNEL SPACING

soft key and

enter the desired channel spacing. For an IS-95 signal, enter 885 or 1.98 MHz.

Step 12. Make the measurement by pressing the

MEASURE

soft key. The Site Master

will display the measurement results in the display area.

The table below provides typical settings for measuring a CW and a IS-95 CDMA signal.

5-5

Chapter 5 Spectrum Analyzer Measurements

CW Wave

IS-95 CDMA Signal

Reference Level

–15 dBm

–10 dBm

Input Attenuation Level

Auto

Auto

Resolution Bandwidth

Auto

10 kHz

Video Bandwidth

Auto

30 kHz

Table 5-2.

ACPR Measurement Settings