What display symbols mean – A&D Blood Pressure Monitor UA-787 User Manual

Page 6

Advertising
background image

LifeSource Health Line (Toll-Free): 1-888-726-9966

E-2

WHAT DISPLAY SYMBOLS MEAN

?

?

Display
Symbol

Recommended
Action

Measurement is in progress,

remain as still as possible.

Take measurement again and

consult with your physician.

Replace all batteries with new

ones when the symbol blinks.

Alkaline batteries recommended.

Try the measurement again.

Remain very still during the

measurement.
Fasten the cuff correctly, and try

the measurement again.

Check for air leaks along the

tube and around the

air socket.
Refasten the cuff and retake

measurement.
Make sure tube is properly con-

nected to cuff and monitor.
Try the measurement again.

Remain very still during

the measurement.
Measurement is in progress,

remain as still as possible.
Refer to the section “About

Pressure Rating Indicator” for

further explanation.

Condition/Cause

The symbol shown while meas-

urement is in progress. It blinks

while detecting the pulse.
An irregular heartbeat or body

movement may have occurred.

Refer to page E-11 for more infor-

mation on irregular heartbeats.
Previous measurement stored in

MEMORY.
The battery power indicator dur-

ing measurement.

The batteries are low when it

blinks.
Unstable blood pressure due to

excessive body movement.

The systolic and diastolic values

are within 10 mmHg of each

other.
The pressure value did not

increase during inflation.

The cuff is not fastened correctly.

There is an air leak in the cuff or

monitor.
The pulse is not detected correctly.

Cuff Inflation Meter

Pressure Rating Indicator™

Battery Full

Battery Low

PULSE DISPLAY ERROR