MAINTENANCE SERVICE AND ANALYSIS GUIDE

MALFUNCTION

POSSIBLE CAUSE

SOLUTION

Refrigerator does not

1. Refrigerator plugged into a

1. Use another circuit.

run.

ground fault interrupt circuit.

2. Temperature control is in the

2. See Temperature Control Section.

OFF

position.

3. Fuse blown or tripped circuit

3. Check/replace fuse with a 15A time delay

breaker.

fuse. Reset circuit breaker

Refrigerator runs too much

1. Refrigerator recently dis-

1. 4 hours required for refrigerator cool down.

or too long.

connected for a lengthy period.

2. Large amount of product

2. Warm product will cause refrigerator to run

recently stored.

more until desired temperature is reached.

3. Doors opened too frequently

3. Open doors less often.

or too long.

4. Refrigerator door may be

4. Refrigerator may not be level.

slightly

open.

5. Temperature control set too low. 5. Reset temperature control.
6. Refrigerator gasket is dirty, worn, 6. Clean or replace gasket.

cracked or poorly fitted.


Vibrating or rattling noise.

1. Refrigerator not level.

1. Re-level the refrigerator as specified in the

INSTALLATION section.

2. Refrigerator is touching the wall. 2. Move the refrigerator away from the wall.


Moisture forms on refrigerator

1. Weather is hot and humid.

1. This is normal.

interior.

2. Door not seating properly.

2. Re-level the refrigerator.

3. Door kept open too long or too

3. Open door less often.

frequently.


Moisture forms on refrigerator

1. Door may not be sealing

1. Re-level the refrigerator.

exterior.

properly.


Door will not close.

1. Refrigerator is not level.

1. Re-level the refrigerator.


Light bulb will not turn on.

1. Light bulb burned out.

1. Replace light bulb with a new bulb of the

Same

wattage.

2. No electric current reaching the 2. Ensure refrigerator is plugged into power

refrigerator.

receptacle.

3. Defective light bulb holder or

3. Contact a service technician.

Light

switch.