GENERAL

INSPECTION

When the equipment is received, all items should be carefully checked against the bill of lading to
insure all crates and cartons have been received. All units should be inspected for concealed
damage by uncrating the units immediately. If any damage is found, it should be reported to the
carrier at once, and a claim should be filed with the carrier. This equipment has been inspected and
tested in the manufacturing facility and has been crated in accordance with transportation rules and
guidelines. Manufacturer is not responsible for freight loss or damage.

LOCATION

The cabinet should also be leveled when it is placed in its permanent location. Do not stack items
on top of the unit. Vibration during shipping and handling may loosen mechanical connections.
Check all connections during installation. Check all wiring and fasteners.

WARNINGS AND CAUTIONS

Warning: Burn Hazard
Do Not
exceed 150°F for non-vented closures; (screw caps, crimp seals, plastic pouches, etc.) Do
Not exceed pre-sterile solution manufacturer’s temperature requirements.
Do Not raise set temperature to increase rate of heating. Allow approximately 4-6 hours for
solutions to reach desired temperatures.
Do Not use liquids on or inject in living tissue unless actual liquid temperature has been measured
and is acceptable. Temperature of warming cabinet’s contents may be hotter than the displayed air
temperature. For patient safety, always check liquid temperature prior to using.

Warning: Electric Shock Hazard.
Do Not remove top electrical cover. Contact a qualified service representative.

Warning: Explosion Hazard.
Do Not use in the presence of flammable anesthetics. Do not heat liquids in the presence of
flammable solvents.

Observe all Warning Labels. Disconnect power supply to eliminate injury from electrical
shock or moving parts when servicing equipment.

INSTALLATION

Door Alignment - If for some reason the doors are not squared up on the cabinet, the doors can be
adjusted. Opening the door(s) and loosening the screws that hold both the top and bottom hinges to
the cabinet can accomplish this. After adjusting the door so that it is aligned correctly, tighten the
screws to securely hold the hinges in place.

Shelving Installation - Locate shelves inside cabinet, install shelf supports (2 RH support, 2 LH
support per shelf.)