SETUP


Pressing the Prg button and holding it down for 5 seconds will display the following setup menu.

SETUP MENU

SET POINTS >

PARAMETERS >







Pressing the  key will move the cursor to the SET POINTS field. Pressing the down arrow key will
display the following screens.

Note: As detailed above, a password of 14 is required to enter the following screens.

Screen 1: Temperature Set Point

TEMPERATURE

SET POINT

160.0

°F








Screen 2: Air Temperature Alarm Set Points

AIR TEMP ALARM

OFFSET>10.0

°F

HIGH:170.0

°F







OFFSET: Offset from the air temperature set point which an alarm will occur.
HIGH: High temperature alarm setting.
LOW: Low temperature alarm setting.

Screen: Air Temperature Alarm Set Points 2

AIR TEMP ALARM 2

ALARM DELAY>120 sec







ALARM DELAY: Amount of time in seconds that an air temperature alarm setting is exceeded
before an alarm will occur.
Pressing the down arrow key while in the PARAMETERS field will display the following screens.




7

2/11 Rev. A 143378