Screen 5: Communication Parameters

COMMUNICATIONS
UNIT IDENT> 1

BAUD RATE> 19200

PROTOCOL>MODBUS







UNIT IDENT: Sets the unit identification for serial communications.

BAUD RATE: Sets the Baud Rate for the serial communications. Baud rates supported; 1200,

2400, 4800, 9600, & 19200.

PROTOCOL:

SUP RS232--For use with Remote Supervisor. Requires optional software,

software key, and RS232 communications board.

SUP RS485--For use with Local Supervisor. Requires optional software,

software key and RS485 communications board.

MODBUS--For use with custom software. Requires optional RS485 communications board.

Screen 6: Analog Output for Air Temperature Re-Transmit

ANALOG OUTPUT SETUP

LOW SETUP>2V

2V> 60.0

°F

10V>180.0

°F

LOW SETUP: This can be set to 2V for 2 to 10VDC output or 0V for a 0 to 10VDC output.

2V(0V): This calibrates the temperature to the lowest voltage output.

10V: This calibrates the temperature to the highest voltage output.