Myron L 6PFCE and 4P User Manual

Page 12

Advertising
background image

b. RES Key

A press of

displays resistivity with units on the right. On the left

is shown solution type selected for resistivity (ref. Solution Selection, pg.

14). The range of display of resistivity is limited to between 10 kilohms

(KΩ) and 30 megohms (MΩ). A solution outside that range will only show

[- - - -] in the display.

c. TDS Key

A press of

displays Total Dissolved Solids with units on the right.

This is a display of the concentration of material calculated from

compensated conductivity using the characteristics of a known material.

On the left is shown solution type selected for TDS (ref. Solution

Selection, pg. 14).

3. pH and ORP/Fr Chl Keys (6PFC

E

)

Measurements are made on solution held in the pH/ORP sensor well

(ref. pH and ORP, pg. 44). The protective cap is removed and the sensor

well is filled and rinsed with the sample enough times to completely

replace the storage solution.

After use, the pH/ORP sensor well must be refilled with Myron L Storage

Solution, and the protective cap reinstalled securely (ref. Maintenance

of the pH/ORP Sensor, pg. 9 and Cleaning Sensors, 2. pH/ORP, pg. 34).

a. pH Key (6PFC

E

)

A press of

displays pH readings. No units are displayed on the right.

b. ORP/Fr Chl Key (6PFC

E

)

In ORP mode, a press of

ORP

Fr Chl

displays Oxidation-Reduction

Potential/REDOX reading in millivolts; “mV” is displayed.
When the FC

E

mode is activated, a press of

ORP

Fr Chl

displays the Free

Chlorine Equivalent reading in “ppm” alternating with the FC

E

predictive

ORP reading in “mV”.

4. CAL/MCLR Key

A press of

allows you to enter the calibration mode while

measuring conductivity, TDS or pH. Once in CAL mode, a press of this

key accepts the new value. If no more calibration options follow, the

instrument returns to measuring (ref. Leaving Calibration, pg. 17).
If

is held down for about 3 seconds when the ORP or FC

E

functions are active, CAL mode is not entered. Instead either “OrP” or

Chl” will appear depending on which mode is active. Change modes

by pressing the Up or Down buttons. Press any parameter key to

exit ORP unit preference selection or let the unit time out. ORP unit

preference will be saved.

8