XIII.

CELL ChECK

The cell check verifies the cleanliness of the conductivity/TDS/resistivity

sensor. In normal use the cell may become dirty or coated and require

cleaning. If the display is showing “.00” when the cell cup is dry, the

sensor is probably clean. However, when testing high purity water in

resistivity (“RES”) mode improved accuracy may be desired. No matter

what a manufacturer claims, a sensor can and will become contaminated

or coated and, therefore, require cleaning. A true 4-wire sensor, as in

the Ultrameter II, helps to mitigate contamination, but NO SENSOR IS

100% IMMUNE.

1.

Press

.

Figure 21

2.

Press

to display the

stored memory records.

3.

Press

repeatedly until

Figure 22

you pass the FAC SEL

location. The display will

show a “CELL ch

(see Figure 21).

4.

Press

to test.

If cell is clean, “Good” will

momentarily be displayed

Figure 23

(see Figure 22). If cell is

dirty, “CELL cLn” will be

displayed (see Figure 23)

(ref. Cleaning Sensors,

pg. 34).

XIV. AUTO oFF

Auto oFF allows the user to adjust the

Figure 24

time the instrument is ON (up to 75

seconds) after each press of a key.

Default time is 15 seconds with 60

seconds in CAL (calibration) mode.

1.

Press

.