solution to flow into and out of the sensor well and cell cup for at

least 10 seconds.

4.

Let sample flow continuously into conductivity cell with no

aeration.

5.

Allow both the sensor well and cell cup to remain filled with

sample.

6.

Press

. The instrument will begin alternating between a

predicted final ORP value and a free chlorine equivalent

concentration in ppm. Both readings will change rapidly at first.

7.

Wait for the readings to stabilize. When the mV and ppm values

are unchanging for 5 consecutive readings, the FC

E

reading

has reached a stable level. This may take 1 to 2 minutes.

nOte: If the reading takes more than 1 minute to stabilize,

press the

after 1 minute to prevent Auto oFF feature

from disturbing the measurement process. Annunciators will

alert you when either the pH or ORP of the final FC

E

ppm value

are Out of Range (“-Or-”).

8.

Press ms to store reading in memory if desired.

3. FC

E

Immersion Method

nOte: Use this method for pools, spas and other large standing bodies

of water.

1.

Hold instrument beneath the surface of the water to avoid

surface effects on the water’s chemistry.

2.

Swirl the instrument around for at least 10 seconds to thoroughly

rinse the cell cup and sensor well.

3.

Continue holding the instrument under the surface while taking

the reading.

4.

Press

.

5.

The instrument will begin alternating between a predicted final

ORP value and a free chlorine equivalent concentration in ppm.

Both readings will change rapidly at first.

6.

Wait for the readings to stabilize. When the mV and ppm values

are unchanging for 5 consecutive readings, the FC

E

reading

has reached a stable level. This may take 1 to 2 minutes.