Myron L 9PTK with FCE Free Chlorine User Manual

Page 33

Advertising
background image

29

calibrate beyond its calibration limits. For all other functions, attempts

to calibrate too far up or down from the ideal “Fac” value stored in the

unit by the factory will cause the displayed value to be replaced with

Fac”. If you accept it (press the cal key), you will have the original

default factory calibration for this measurement. The need to calibrate

so far out that “Fac” appears indicates a procedural problem, incorrect

standard solution, a very dirty cell cup or an aging pH/ORP sensor (ref.

Troubleshooting Chart, pg. 54).

C. Calibration Procedures

1. Conductivity or TDS Calibration

a.

Rinse conductivity cell 3 times with proper standard (KCl, NaCl,

or 442) (ref. Cond/TDS Standard Solutions, pg. 56). For user

calibration see User Calibration Conductivity/TDS below.

b.

Refill conductivity cell with same standard. KCl-7000 shown.

c.

Press

or

, then press

; “cal” icon will

appear on the display (see Figure 5).

d.

Press

or

to step the displayed value toward the

standard’s value (7032 > 7000) or hold a key down to scroll

rapidly through the reading.

e.

Press

once to confirm new value and end the calibration

sequence for this particular solution type. If another solution

type is also to be measured, change solution type now and

repeat this procedure.

2. User Calibration Conductivity/TDS

Instrument must be in User mode, see Solution Selection, pg. 24.
a.

Rinse conductivity cell three times with YOUR standard.

b.

Refill conductivity cell with same standard.

c.

Press

or

, then press

twice in COND/

three times in TDS. The “cal” icon will appear on the display.

Advertising