27

a.

While in User mode, press

.

b.

Press

twice (to skip over tempco adjustment) and

ratiO” will appear (see Figure 4).

c.

Adjust with

or

key until new conversion ratio is

displayed.

d.

Press

twice (to skip over calibration adjustment) to

accept new conversion ratio. You are now ready to measure

samples with the new conductivity/TDS ratio.

vii. calibratiOn

A. Calibration Intervals

generally, calibration is recommended about once per month with

Conductivity or TDS solutions. Calibration with pH solutions should

be checked twice a month. Calibration of ORP is not necessary (ref.

CALIBRATION INTERVALS TABLE under Calibration Steps, pg. 28).

B. Rules for Calibration of the Ultrameter III

1. Calibration Steps

The following table lists measurement functions and their corresponding

calibration requirements: