Checking the volume – BrandTech Scientific HandyStep S User Manual

Page 34

Advertising
background image

Checking the Volume

*) = Calculation of accuracy (A%) and variation coefficient (CV%):
A% and CV% are calculated according to the formulas for
statistical control.

x

i

= Weighing results

n = Number of weighings

Z = Correction factor

(e.g., 1.0029 µl/mg

at 20 °C, 1013 hPa)

V

0

= Nominal volume

Calculation (for nominal volume)

Accuracy*

Coefficient of Variation*

Standard Deviation

V – V

0

V

0

A% = · 100

Σ (x

i

– x)

2

n – 1

s = Z ·

100 s

V

CV% =

Mean value x = Σ x

i

n

Mean volume V = x · Z

Testing instructions (SOPs) and a demo version of the EASYCAL

TM

4.0

calibration software are available for download at www.brand.de.

Note:

34

Advertising