Warranty – Chemglass CG-15001 Series User Manual

Page 10

Advertising
background image

10

one minute, if the light comes on the controller is
functioning properly.

inPt FAiL” flashes

Temperature sensor is unplugged,
corroded, or broken

Ensure Temperature sensor is plugged in. Clean or
replace.

-AL-“ flashes

The process temperature is higher
(+10

C) than the set point

See Over-Temp Protection section, pg. 9.

PArk” flashes

Controller has been placed in
„Park‟ mode

1. Press and hold both the  and 

keys until “tunE” is

displayed. 2. While holding the  key, press the  key
until “oFF” appears. 3. Press and hold the  and  keys
until the temperature is displayed.

tunE” “FAiL” flashes

Autotune routine has failed

Turn off Controller for 10 seconds and retry. If problem
persists, call Chemglass‟ Technical Service.

Warranty

Chemglass, Inc. warranties this unit against defects in material and workmanship for a period of one year from the date of sale. If
the unit should malfunction it must be returned for evaluation. If the unit is determined to have a defect in materials or
workmanship then it will be repaired or replaced at no charge. Tampering or damage resulting from excessive current, heat,
moisture, vibration, corrosive materials, or misuse will void this warranty. Programming changes or reconfigurations are not
covered under warranty. To return this item under warranty, advise Chemglass technical service department of the situation and
they will provide you with a return authorization number. The unit will be evaluated and repaired as required in accordance to
the warranty.

Please contact the Chemglass technical service department for thermocouples and accessories along

with warranty repair information.

Chemglass technical service department:

tel: 1-800-843-1794 ext. 2124

fax: 1-800-922-4361