Chemglass CG-15001 Series User Manual

Page 3

Advertising
background image

3

Q

Q

U

U

I

I

C

C

K

K

S

S

E

E

T

T

-

-

U

U

P

P

*This unit has been factory tuned for use with Heating Mantles sized 1L to 12L.

1) Ensure that the Power Switch (On/Off switch) on the Controller is in the OFF position.
2) Plug Power cord of Chemglass 15001 Series Controller into a 120VAC receptacle.
3) Plug the Heating Device Cord into the 120VAC power receptacle on the Controller.
4) Insert the correct Thermocouple Plug into the Thermocouple Receptacle on Chemglass 15001 Series Controller. NOTE:

Color of Thermocouple Plug must be the same color as the Thermocouple Receptacle.

5) Place Thermocouple Probe into medium. NOTE: Probe should not be in contact with bottom or sides of heating vessel,

and should be submerged 2/3rds of medium content.

6)

Turn Power Switch “ON”, and then enter the desired temperature set point via the Controller‟s keypad. Press and hold
the  key, then press  key to increase or, press  key to decrease set point. Once set, the Controller will display the
actual temperature of the medium being measured. The set point can be displayed at anytime by pressing the  key.

FRONT PANEL VIEW


ON


OFF


T


1500

1

2

3

4

5

6

7

8

Status of Heating

Element

Control Key ()

Down Key ()

Up Key ()

Power (ON/OFF)

Switch

DISPLAY (Set point,
temp, etc.)

Power Receptacle
(120VAC)

Thermocouple
Receptacle