4

Operator’s Manual for the Bullet Blender

®

50

minutes. To operate, press the “START”
button. After 1 minute, the LED light on
the number 1 will light up. After 2
minutes the second LED light by the
number 2 will light up. And so on. At the
setting shown above, after 12 minutes,
the Bullet Blender

®

50 will stop.

The BB50-AU model has a “4°C”

light that comes on when the internal
compartment of the instrument is
refrigerated. On the BB50-DX model, the
“Air Cooling™” light indicates that
ambient air is circulating through the
inside of the instrument during operation
and for 30 seconds after the cycle is
complete.

The BB50-AU unit contains a dry ice

cooling compartment. To use, fill the dry
ice cooling compartment with at least 2
lbs of dry ice. Always use cryogenic
gloves when handling dry ice. Do not use
the inner portion of the dry ice cooling
compartment to transport the ice. Do not
fill the dry ice cooling compartment with
water ice. Make sure that the dry ice
cooling compartment lid is properly
sealed. To pre-cool the instrument, after
filling the compartment with dry ice, run
the instrument for 5 minutes.

PROTOCOLS AND SAMPLE

SETTINGS

The following ratio should be used as

a guideline for determining the amount of
beads and buffer to use given a certain
sample size - 1 volume/mass of tissue : 1
volume of beads : 2 volumes of buffer.

For more specific information regarding the

use of various beads as well as specific

protocol information, please refer to our

website: www.nextadvance.com.

For 50mL tubes, do not operate with

more than 3.5g of sample, or a total of 20mL

of buffer, tissue, and beads (per tube).

Cutting the tissue into smaller pieces

will generally yield better results. Tissue

with a high aspect ratio (long, thin strips)

will homogenize better than tissue that is

round or cubic.

Do not operate the Bullet Blender

®

50

using the same tubes for more than 30

minutes.

Protocols for many types of samples are

posted

on

our

website,

at

www.nextadvance.com/FAQs/protocols.htm.

Notes:

Use Axygen

®

brand or Corning

®

brand

polypropylene, plug seal cap, skirted (self-

standing) 50 mL tubes. If there is a different

brand of 50 mL tube you would like to use,

please

contact

us

at

[email protected]. There may be

some discoloration of the sleeves and of the

tubes after homogenization. This is normal

and a by product of the workings of the
Bullet Blender

®

50, which is designed to

impact the tubes with as much energy as

possible.

CLEANING

If you wish to clean your Bullet Blender,

To use Air cooling

, where ambient

air is drawn into the instrument, operate

the BB50-AU unit with the dry ice

compartment open. Air Cooling

is most

effective when the instrument is operated

in a 4°C environment.