SP1600/1800 Programmable Syringe Pump

Publication #1200-02

16

9/15/04

7.2 Audible Alarm Enable

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When alarms are enabled, the buzzer will be sounded as follows:

Condition

Buzzer Action

Pumping Program ended

Continuous beeping

Pumping Program paused for start trigger

Continuous beeping

Alarm condition, such as pump motor stalled

Steady alarm

Pressing any key, will stop the alarm.

7.3 TTL I/O Operational Trigger Configuration

aa

Configures how the TTL I/O ‘Operational Trigger’ (pin 2) will control the Pumping

Program’s operation. (See sec. 10.1, TTL I/O Operational Controls). The 2 letter configuration parameter
to the right of the colon (:) is defined as follows:

Setting

Name

Function

Ft

Foot Switch

Falling edge starts or stops the Pumping Program

LE

Level Control

Falling edge stops the Pumping Program

Rising edge starts the Pumping Program

St

Start Only

Falling edge starts the Pumping Program

Sp

Stop Only

Falling edge stops the Pumping Program

7.4 TTL I/O Directional Control Input Configuration

aa

Configures how the TTL input ‘Pumping Direction’ (pin 3) will control the pumping

direction. (See sec. 10.1, TTL I/O Operational Controls). The 2 letter configuration parameter to the right
of the colon (:) is defined as follows:

Setting

Name

Function

rE

Reciprocating Pumps

Falling edge: Infuse; Rising edge: Withdraw

dU

Dual Pump

Falling edge: Withdraw; Rising edge: Infuse

7.5 Pump Motor Operating TTL Output Configuration

n

Configures the functionality of the 'Pump Motor Operating' TTL output pin (TTL pin 7).

Settings: 0: Sets the output to logic high only when the motor is operating (pumping).

Sets the output to logic low when the motor is not operating or when the Pumping Program is

executing a pause timer or is stopped

1: Sets the output to logic high when the motor is operating (pumping) or when the Pumping

Program is executing a pause timer.
Set the output to logic low when the Pumping Program is stopped