42

9/15/04

Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.

<n> Indicates pin number on ‘TTL I/O’ connector

Valid value: 5 (Program Output pin)

Query is not applicable.

TTL INPUT QUERY

IN <n>

Queries TTL level of pin on ‘TTL I/O’ connector. Set is not applicable.

<n> Indicates pin number on ‘TTL I/O’ connector

Valid values: 2, 3, 4, and 6.

Response: <TTL level>

BUZZER

BUZ [ 0 | { 1 [ < n > ] } ]

Sets / queries buzzer

Set:

0 = Turn buzzer off;

1 = Turn buzzer on

if <n> specified

If <n> = 0, buzzer beeps continuously,
otherwise buzzer beeps <n> times

if <n> not specified, buzzer sounds continuously

Query response: { 0 | 1 }

0 = Buzzer off
1 = Buzzer is on continuously or beeping.

FIRMWARE VERSION QUERY

VER

Response: NE100MV <n> . <n>

where ‘<n>.<n>‘ is current firmware version number.

Set is not applicable.

MASTER PROGRAM RESET

* RESET

Clears program memory and resets communication parameters to Basic mode and address 0.

This is a special system command that will be accepted by the pump regardless of its current address.

9.5 Getting Started With RS-232

Before beginning to develop pump control software for a computer, first setup and experiment with the
pump’s communication. After attaching the pump to the computer, run a terminal emulation Program on
the computer. A generic terminal emulator, supplied as standard software with many computers, can be
used to communicate with the pump in the Basic communications mode.

Next Advance supplies a terminal emulator, for demonstration purposes only, which allows more complex
control of the pump. Also, this terminal emulator contains a “Pump Programming Language” (PPL™),
which allows Pumping Programs to be developed using symbolic text, modified, and stored in computer
files, then downloaded to the pump.

With a generic terminal emulator, setup the terminal emulator with the same baud rate as the pump and
with an 8 bit data, no parity, and 1 stop bit (8N1) data frame. Set the communications port to the port that
is attached to the pump. Also enable local echo (half-duplex) and turn flow control off.