Appendix, 1 syringe diameters and rate limits – Next Advance SP1800 User Manual

Page 52

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1600/1800 Programmable Syringe Pump

Publication #1200-02

47

9/15/04

11. Appendix

11.1 Syringe Diameters and Rate Limits

Manufacturer

Syringe

(cc)

Inside

Diameter
(mm)

Maximum

Rate
(mL/hr)

Minimum

Rate
(µl/hr)

B-D

1

4.699

34.67

0.454

3

8.585

115.7

1.515

5

11.99

225.7

2.954

10

14.43

327

4.279

20

19.05

569.9

7.457

30

21.59

732

9.578

60

26.59

1110

14.53

Monoject

1

5.74

51.74

0.677

3

8.941

125.5

1.643

6

12.7

253.3

3.314

12

15.72

388

5.078

20

20.12

635.7

8.318

35

23.52

868.7

11.37

60

26.64

1114

14.59

Terumo

1

4.7

34.69

0.454

3

8.95

125.7

1.646

5

13

265.4

3.473

10

15.8

392

5.13

20

20.15

637.6

8.343

30

23.1

838

10.97

60

29.7

1385

18.13

Air-Tite

10

15.9

397

5.195

20

20.25

643.9

8.426

30

22.5

795

10.41

50

29

1320

17.28

Syringe
(uL)

Inside
Diameter

(mm)

Maximum
Rate

(ul/hr)

Minimum
Rate

(µl/hr)

Hamilton
Microliter

0.5

0.103

16.66

0.001

1

0.146

33.47

0.001

2

0.206

66.64

0.001

5

0.326

166.9

0.003

SGE

0.5

0.1

15.7

0.001

1

0.15

35.33

0.001