PS5000C MKII Operator’s Manual

3-10

2.2

Aging - 10min

This adjustment is to stabilize the lamp and the CCD.

NOTE !

It will take time until the lamp is stable especially in the morning in
winter. In such a case, repeat the “Aging – 10min” procedure about
two times.

1

Launch the application and the PS5000C MKII TWAIN Driver.

2

Check “Adjustment” on the screen of the PS5000C MKII TWAIN Driver.

3

Select “1. Aging - 10min” on the pull-down list under “Adjustment”, and then
click “Start”.

4

Click “OK” on the confirmation screen.

5

The window indicating that adjustment is underway will appear. When the
window disappears, the adjustment is completed. Proceed to the next
adjustment, “2.3 Exposure Time”.