BK Precision 4070A User Manual Rev.2.2

41


You now have a terminal connection established with the 4070A. Hit the “?” key. You should see a command
menu in response.
If you do not see this menu, check the following:

On the 4070A, press Mode then Offset and verify that the baud rate is set to 9600.

Verify that you are using the correct COM port.

Check your cabling.

If you see the command menu, you have verified that the 4070A is properly cabled to the computer, and that you
are certain which COM port you are using.

NOTE: In general any terminal software capable of sending standard ASCII codes over RS-232 terminal could be
used to control 4070A remotely, please refer to software provider for specific terminal OS/software support issues.

6.4 Remote Control Operation

Each key on the front panel keypad of the 4070A has an associated ASCII character which, when sent to the
4070A over the serial port, has the same effect as pressing that key on the keypad.
At power-up, the 4070A RS-232 baud rate is recalled from stored configuration 0. The factory default for this is
9600 baud. The remaining serial port parameters are always 1 start bit, 8 data bits, 1 stop bit, and no parity. The
serial port on the controlling computer must be set to match these values.

6.5 Remote Control Programming Rules

The following guidelines should be observed when writing control software for the 4070A:

All ASCII commands are case insensitive, i.e. upper and lower case letters are treated equally by the
4070A.

When the 4070A has finished executing a command, it will return a command prompt, which is the
">"character. The control software should clear its receive buffer, issue a command, and then look for the
">" character to determine when the command has been executed. If a long string of commands is sent to
the 4070A, a separate ">" character will be returned for each command as it is executed.

All commands require less than 300 mS to execute except for the store or recall commands, which require
several seconds to complete.

All white space (carriage returns, linefeeds, tabs, spaces and commas) between commands is ignored.
Invalid commands (ASCII characters that are not listed in the command menu) are likewise ignored.

A command does not have to finish execution before another can be sent. All characters received by the
4070A are buffered up and executed in the order they were received. The programmer should allow
sufficient time for the 4070A to execute each command and for the signal output to settle into the desired
state.

If the 4070A is reporting data to the control program, it will place a colon (:) character before the data. This
makes it easy for the control program to parse a returned string and extract the character(s) needed. Some
examples of returned values are:


Example 1
DTMF Detection Mode detected digits:
DTMF: #

Example 2 Hardware/Software version, serial number (the "V" command):

BK Precision model: 4070A
Software Version: c.2
Hardware Version: 1.0
S/N: F45E3412AC56
PM Checksum: 0017829BB903

When the operating mode of the 4070A is changed, all parameter values for the previous mode are retained in
memory as long as power is applied to the unit. For example, if Sinewave mode with an output frequency of 5 MHz
is changed to sweep mode, the Sinewave mode 5 MHz value will be saved before the mode is changed. Switching
back to Sinewave mode will restore the 5 MHz value in the frequency field. If power is turned off, however, all
parameters are reset to default values. To save all parameters for a mode, use the "Store/Recall" command.