9 multiple units locking, Master arb generator, Slave arb generator – B&K 4070A User Manual

Page 65: 1 introduction to multiple arb locking, Quick start guide, How it works, Multi unit hookup

Advertising
background image

BK Precision 4070A User Manual Rev.2.2

60

7.9 Multiple Units Locking

7.9.1 Introduction to Multiple Arb Locking

With this feature, users may lock several arbitrary waveform generators together to generate multiple simultaneous
signals, which are locked together in frequency. The user may then adjust a phase-offset field to obtain a desired
phase relationship between the waveforms with a minimum phase resolution of 0.01 degrees. Any shape of
waveform may be used. In fact, waveforms of different shapes may be locked together as long as they all have the
same number of points.

Quick Start Guide

To implement Multiple Unit Locking, follow these steps:

1. Cable units together as shown below.
2. Place the Master Unit in Arbitrary Waveform Mode and set the mode on the LCD display to Lock Master
3. Likewise set the mode on each slave unit to Lock Slave
4. Load waveforms into the Master and Slave units. The waveforms may differ in shape but they must each

have the same number of points. To generate phase offsets up to 359.99 degrees, the number of points must
be either = 32,768 or

16,384 (refer to the following section “Waveform Size Restrictions”).

5. Set the sampling rate on the Master Unit to the desired sampling frequency. (The frequency shown on the

slave units is not used).

6. If desired, set the phase offset on each Slave Unit(s) to the desired phase offset(s).
7. Optionally save the setup on each unit using the Store/Recall button. The waveforms will also be saved.

How it works

Arb wave locking works by using common timing signals for multiple arbitrary waveform generators. To lock
multiple generators together, two signals are shared: a sample clock and a reset signal (see the diagram below). The
sample clock ensures that the waveform points on all generators are generated at exactly the same rate. The reset
signal tells all units to “jump to the first sample point” at the same time. With these two signals, all units are
guaranteed to be on the same sample point at the same time. The 4070A uses one arbitrary waveform generator as a
“Master” unit that generates the sample clock and reset signal for all of the “Slave” units.

Multi Unit Hookup

Two cables are used to connect on master 4070A to one or more slave units.

NOTE: Observe that one of the connectors for the interconnecting cables is on the Front Panel. You may hook
from two to six units together this way.

Master

Arb

Generator

SYNC

Out

(front panel)

SIG
Out

(front panel)

ARB

Sync I/O

(rear panel)

Slave

Arb

Generator

Ext Arb

Clk In

(rear panel)

SIG
Out

(front panel)

ARB

Sync I/O

(rear panel)

Slave

Arb

Generator

Ext Arb

Clk In

(rear panel)

SIG

Out

(front panel)

ARB

Sync I/O

(rear panel)

to
more
units

Sample Clock

Reset Sig

Waveform

#1

Waveform

#2

Waveform

#3