BK Precision 4070A User Manual Rev.2.2

69

8.0 DC Operation Option

Description

The DC Operation Option allows the user to power the 4070A from a DC voltage source in the 9-36 VDC range.
The user may easily switch between the AC Line supply and a DC source. This option is most useful for service and
remote applications where AC power is not available.
A 20-72V input voltage range is also available. Please contact BK Precision for availability.

Specifications:

Input Voltage Range: 9-36 VDC

Max. Power Consumption: 15 W

Switching Frequency: 200 KHz

Isolation Voltage: 500 VDC


Wiring Diagram

+ 9-36 VDC

To Power
Input

DC to DC
Converter

Red

Black


*** Caution***

Be certain to observe proper polarity of the input voltage connection. Permanent damage may

result from reversal.