background image

Normal Control Solution

High Control Solution

Accuracy

1. Bayer Accuracy Evaluation:

To assess the accuracy of the Ascensia

®

B

REEZE

™ System,

102 fresh fingerstick specimens from people with diabetes
were tested with the Ascensia

®

B

REEZE

™ (2 readings per

fingerstick) and the YSI 2300 STAT Plus Glucose Analyzer.
The glucose concentration of the samples ranged from 2.3
to 30.1 mmol/L with an average of 9.9 mmol/L. The range
of hematocrits was 29% to 55% with an average of 40%.

33

Total

Total

Strip No.

of

Mean

SD

CV

Lot

Operator

Assays

(mmol/L) (mmol/L)

(%)

A

Lay User (n=50)

100

8.394

0.400

4.8

B

Lay User (n=50)

100

8.317

0.356

4.3

Total

Total

Strip No.

of

Mean

SD

CV

Lot

Operator

Assays

(mmol/L) (mmol/L)

(%)

A

Lay User (n=50)

100

18.706

0.961

5.1

B

Lay User (n=50)

100

18.867

0.800

4.2