Comparison of meter to laboratory results, B. healthcare professional results – Bayer HealthCare Ascensia BREEZE and Ascensia AUTODISCTM Unique 10-Test Disc User Manual

Page 42

Advertising
background image

Comparison of Meter to Laboratory Results

† Weighted Deming regression, weighted linear correlation coefficient
Opr. = Operator; Lay = Lay User; HCP = Healthcare Professional

36

0.0

5.0

10.0

15.0

20.0

25.0

30.0

35.0

0.0

5.0

10.0 15.0 20.0 25.0 30.0 35.0

Laboratory Plasma Glucose Result (

mmol/L)

Ascensia

®

B

REEZE

™ Result (mmol/L)

o = Lot A

+ = Lot B

y=x

95% Confidence Intervals

Lot

Opr.

N

Y=

Slope

/

Intercept

R

A

Lay

100

0.85x + 0.72

0.81

to

0.89

0.38

to

1.06

0.98

HCP

100

0.84x + 0.65

0.80

to

0.88

0.33

to

0.96

0.98

B

Lay

96

1.02x – 0.58

0.96

to

1.08

–1.04

to

–0.13

0.97

HCP

96

1.03x – 0.61

0.99

to

1.08

–0.94

to

–0.29

0.98

A&B

All

392

0.91x + 0.21

0.89

to

0.94

0.01

to

0.40

0.97

B. Healthcare Professional Results