Bayer HealthCare Ascensia BREEZE and Ascensia AUTODISCTM Unique 10-Test Disc User Manual

Page 48

Advertising
background image

42

What You See

What It Means

What You Do

Your Blood Glucose Level is
very high or meter error.

Turn meter off, then on
using

. Retest. If error

appears again, call your
physician or healthcare
professional (HCP). Your
blood glucose may be
above 55.5 mmol/L (1500
mg/dL). If the problem is
with the meter and cannot
be corrected, contact the
Bayer Diabetes Helpline
(see pg. 45).

There was a problem filling
the test strip completely
full.

Review pg. 8, and repeat
test.

Meter has detected an
error; accurate test result
not possible.

• You may have applied

blood too early to the test
strip. Repeat test and wait
until you see the flashing
blood drop to apply blood.

• There might also be

moisture on the test strip
guide. Clean meter if
needed and dry
thoroughly (see pg. 27).
Repeat test.

Meter has detected an
error; accurate test result
not possible.

• Test strip is not

completely filled. Release
used test strip and repeat
test. (see pg. 8) or

• 10-test disc may have been

damaged due to intense
heat or moisture exposure.
Replace 10-test disc, using
a new box if necessary.

Advertising