Bayer Esprit User Manual

Page 20

Advertising
background image

CHAPTER

Problem Solving / Warning Codes

14

31

CODE

CAUSE / SOLUTION

Temperature is too low (between
0° and 10° C) and could affect test
results.
■ Test within operating temperature

range (10° to 40° C).

Temperature is too high
(between 40° to 50° C) and could
affect test results.
■ Test within operating temperature

range (10° to 40° C).

Low Batteries
■ Replace batteries within one

week or 20 readings.

30

CHAPTER

14

Problem Solving / Warning Codes

CODE

CAUSE / SOLUTION

Test Result is below 0.6 mmol/L.
■ Meter has detected an error.

Repeat the test and make sure
Test Sensor is filled.

Test Result is above 33.3 mmol/L.
■ Repeat the test. If still high,

contact your physician or diabetes
educator regarding a possible
hyperglycemic condition (high
blood sugar).

mmol

L

mmol

L

Deleted Test Result
■ If the temperature is too low or

too high, the test result is marked
with an X and deleted from the
average.

mmol

L

EXAMPLE