CHAPTER

Technical Information

15

37

Glucometer

®

E

SPRIT

™ System

(Whole Blood Referenced)

1. Accuracy: A patient-use clinical study was
performed at diabetes care centers in three
different locations. Glucose levels were measured
with the Glucometer E

SPRIT

System on over

300 fresh capillary blood specimens by 61 persons
with diabetes, and six healthcare professionals.
The results were compared with the whole blood
glucose results of the YSI 2300 Stat Plus (Yellow
Springs Instrument Co., Inc., Yellow Springs, OH),
a glucose oxidase/electrode system.

Figure 1 on the following page shows the
comparison of the lay diabetic Glucometer
E

SPRIT

System results with the whole blood

glucose results. The comparison of the healthcare
professional results with the whole blood glucose
results is shown in

Figure 2.

36

CHAPTER

15

Technical Information

REPLACEMENT PRODUCTS:

When calling or writing for supplies, be sure to include the
number with the name of the replacement part or product.

Part Number

Item

40030032

3-volt Lithium Battery (DL or
CR 2016) — 2 (Available at most stores)

99939679

Glucometer

®

E

SPRIT

™ User Guide

50184147

Glucometer

®

E

SPRIT

™ Quick Reference Guide

Part Numbers subject to change without notice.

Product
Number

Product Name

3616

Glucometer

®

E

SPRIT

™ Test Sensor Discs — 50s

3649

Glucometer

®

E

SPRIT

™ Low Control

3650

Glucometer

®

E

SPRIT

™ Normal Control

3651

Glucometer

®

E

SPRIT

™ High Control

6541

Microlet™ Automatic Lancing Device

5970

Glucolanz™ Lancets (200s)

6555

Microlet™ Endcaps — Regular (10s)

6559

Microlet™ Endcaps — Super (10s)

Replacement products may be obtained from Diabetes
Care Centers, or your local pharmacy.