Instruction 24-9448

5-4

Calibration

5.4 T-Stack Calibration

Ths procedure frst zeros and then spans the stack-temperature channel

to known temperature values.

The use of an electronc thermocouple smulator s the preferred method of

producng the desred calbraton temperatures. Alternatvely, contaners

of ce water and bolng water can be used.

Material Required:

Thermocouple Smulator (K-type)

Range: 0 to 600 °F

Accuracy: ±0.5 °F

(Alternatvely) Ice-Water, Bolng Water, Thermometer

TS-Zero Procedure:
1. Set thermocouple smulator to room temperature and plug ts output

nto the T-STACK connector located at the bottom of the analyzer.

Alternatively: Plug the probe’s thermocouple nto the T-STACK con-

nector located at the bottom of the analyzer. DO NOT attach the

probe’s gas hose to the analyzer’s GAS port; otherwise water

will be drawn into the analyzer!

2. If not already done, turn ON the analyzer and dsplay the

CALIBRATION LIST screen per Secton 5.2.

3. Use the buttons to hghlght T-Stack, and

then press

ENT

to dsplay the CALIBRATE

TS-ZERO screen.

“Measured” is the current temperature read-

ing, while “Applied” is a known temperature

that will be applied for calibration purposes.

4. Set thermocouple smulator to 32 °F (0 °C),

and then use the and buttons to enter

an Appled value that exactly equals the set-

tng of the smulator.

Alternatively: Submerge probe tp nto an ce-water bath wth a ther-

mometer, wat several mnutes, and then use the and buttons to

enter an Appled value that exactly equals the thermometer readng.

The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to

calibrate outside this range will cause the message “Bad Calibration

Wrong CAL Entry” to appear in the following step.

-

-

CALIBRATION LIST

Pressure

T-Stack

T-Air
CO-LO
SO

2

NO
NO

2

(more)