Barco MDRC-2124 User Manual

Page 8

Advertising
background image

8

Intended use

This display is an AMLCD display designed for viewing medical X-ray
images. This unit should not be used near patients and should be kept
outside of 1.83 m perimeter and 2.29 m vertical.

Power connection

Power requirements: The display must be powered using a medical
approved 24VDC power supply.

For use only with power supply SINPRO:HPU100-108
Input:100-240V~, 47-63Hz, 1.2-0.5A

The medical approved DC power supply must be powered by the AC
mains voltage.

Power cords

Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent,
18 AWG min., rated 300 V min., provided with a hospital-grade type
plug 5-15P configuration for 120V application, or 6-15P for 240V
application.

Do not overload wall outlets and extension cords as this may result in
fire or electric shock.

Mains lead protection (U.S.: Power cord): Power cords should be routed
so that they are not likely to be walked upon or pinched by items
placed upon or against them, paying particular attention to cords at
plugs and receptacles.

External equipment

External equipment intended for connection to signal input/output or
other connectors, shall comply with relevant UL/ EN/ IEC standard (e.g.
UL/EN/IEC 60950 for IT equipment and UL/EN 60601-1 / IEC 60601
series for medical electrical equipment). In addition, all such