ISMATEC SA - Pumpenköpfe/Pump Heads/Têtes de pompe / 11.01.05/CB/GP

10

Only for MCP Process
from version 3.0

Entering the pump-head
k In order to obtain correct pumping
and dispensing values, the individual
ID code of the mounted pump-head
must be entered; especially when
a pump is used for the fi rst time
and each time the pump-head is
changed.

1 Select the program (1 to 4)

2 Switch the pump off

3 Keep the settings key pressed and

switch the pump on

k Pump is now in settings menu

4 Select

HEAd

with the

56

keys

Head = ID code
(Pump-head - Can be set separa-

tely for each program)

Nur für MCP Process
ab Version 3.0

Pumpenkopf programmieren
k Für korrekte Pump- und Dosierwer-
te muss der ID-Code des montierten
Pumpenkopfes eingegeben werden;
vor allem bei der ersten Inbetriebnah-
me und nach jedem Pumpenkopf-
wechsel.

1 Programm wählen (1-4)

2 Pumpe ausschalten

3 Die settings-Taste gedrückt halten

und Pumpe einschalten

k Pumpe ist nun im Settings-Menu

4 Mit den

56

Tasten

HEAd

wählen

Head = ID-Code

(Pumpenkopf - Für jedes Pro-
gramm individuell einstellbar)


Seulement pour MCP Process
à partir de la version 3.0

Séléctionner la tête de pompe
k Pour obtenir des valeurs de
pompage et de dosage correctes,
il faut saisir le code ID de la tête de
pompe installée; ceci surtout lors de
la première mise en service et après
chaque changement de la tête de
pompe.

1 Choisir le programme (1-4)

2 Déclencher la pompe

3 Maintenir la touche settings
enfoncée et enclencher la pompe

k La pompe se trouve dans le
menu des réglages de base

4 Sélectionner

HEAd

avec les

touches

56

Head = Code ID de la tête de pom-

pe (Réglage individuel pour chacun
des 4 programmes)

aus/off
auf Rückwand
on rear panel
sur panneau arrière

2

3

4

settings

+

settings

1

Advertising