SECTION 2

RS-232 COMMANDS

A PC can control all parameters of the model PB-5. The following commands will
allow full operation of the pulser including the RAMP mode. This is convenient when
remote operation of the PB-5 is required (see last paragraph of Section 1.3 for
information on configuring the PC). This full set of commands can be conveniently
displayed by typing "help".

set trigger mode internal

set trigger mode external

set trigger mode gated

set trigger mode one pulse

trigger one pulse

set threshold 3.5

set pulse on 1/0

set amplitude 10.000

set rep rate 10000

set width 25000

set delay 10000

set rise time 10000

set fall time 10000

set tail pulse 1/0

set polarity positive 1/0

set attenuation 1000

set display kev 1/0

set equivalent kev 1000

clamp baseline 0/1

set ramp startv 2.0

set ramp stopv 10.0

set ramp startev 100.0

set ramp stopev 1000.0

set ramp time 100

set ramp cycles 10

exrcute ramp

recall factory defaults

help

Values given for time in the above examples are in nanoseconds.