3.6 COMMAND FORMAT

[Channel]PB5_Command Parameter

on Multi Channel Mode and Single Channel Mode.

PB5_Command Parameter

on Single Channel Mode only.

example (Multi Channel Mode):

1 Frequency 100

will set PB-5 connected on Channel 1 to 100 Hz.

8 Delay 500n

will set PB-5 connected on Channel 8 to 500 ns delay.

5 Load c:\Lab\Set1.bnc

will load settings saved as file Set1.bnc located at
folder c:\lab\ to PB-5 connected on Channel 5.

example (Single Channel Mode):

Width 1µ

will set PB-5 to 1 µs Pulse Width.

Rise 2e-6

will set PB-5 to 2 µs Rise Time.

NOTE:

PB5 Host is programmed to accept common numerical notations in its
parameters.
All of the following are accepted and will produce the same parameter value:

0.0000005, 500n, 500e-9, 0.5e-6, and 5e-7 are all evaluated as 500
nano.

1500 and 1.5K are evaluated as 1.5 kHz when setting freq.

PB5 Host is not case sensitive, thus command 1 TRIG int and 1 TrIg InT will
set channel 1 to internal trigger or internal clock.

3.7 PB5 HOST COMMAND SUMMARY

Command

Keyboard

Shortcut

Parameter Range

Description

FREQuency

Q & Z

2 - 500K

Set frequency to parameter
specified.
Ctrl-Q set 1 Hz Up
Ctrl-Z set 1 Hz Down
Alt-Q set 10 Hz Up
Alt-Z set 10 Hz Down.

WIDTh

W & X

100n - 10m

Set pulse width to parameter
specified.
Ctrl-W set 1 ns* Up
Ctrl-X set 1 ns* Down
Alt-W set 10 ns* Up
Alt-X set 10 ns* Down.
*Varies depending on range.