BNC PB-5 User Manual
Page 16
16
Command Summary (-cont-)
Command
Keyboard
Shortcut
Parameter Range
Description
TRIGger
INternal,
EXternal, GAted,
ONe, FIre
Set trigger to external clock or
external trigger or gated or
single cycle (1 pulse) and Fire
to issue 1 pulse.
THREshold
0
3.5
Set threshold level to
parameter specified.
DISPlay
KEv / VOlt
Set display either keV or
Volt.*
EQUIvalent
100 - 9999
Set keV equivalent voltage.
CLAMp
ON, OFf
Set pulse clamping either on or
off.
RAMP
EXEC
Execute ramp
RAMP
VBEGin
0
9.5
Set ramp start voltage**
RAMP
VEND
0.5 - 10
Set ramp stop voltage**
RAMP
KBEGin
0 - 9499
Set ramp start keV***
RAMP
KEND
5 - 9999
Set ramp stop keV***
RAMP
CYCLes
0 - 999
Set number of Ramp Cycles
RAMP TIME
30 - 900
Set ramp time.
SAVE
Path & Filename Save setting to a file
LOAD
Path & Filename Load saved setting to PB-5
NOTE:
* No Effect on REMOTE Display. Will only affect LOCAL Display.
** Start and Stop Voltage must have a minimum of 0.5 V difference.
*** Minimum and maximum value is affected by equivalent keV settings, the values must be
proportional to the Voltage equivalent.